Activity overview
Latest activity by Jani Timonen-
Jani Timonen created an article,
Problem in Medikro Spirometry Software final report: “TextFileParcer.exe stopped working”
We have found a problem in the final report of Medikro Spirometry Software version 4.0, which causes a situation where the error message “TextFileParcer.exe stopped working” appears and the person’...
-
Jani Timonen created an article,
Bulletin on the use of Medikro's spirometry equipment during a coronavirus (COVID-19) pandemic
The coronavirus pandemic continues to be active around the world. Despite the epidemic, Medikro products are safe to use. Due to our disposable flow sensor and the design of the pressure hose, ther...
-
Jani Timonen created an article,
MSS v4.9 is now released!
OUT NOW Latest Medical Spirometry Software (MSS) Version 4.9 We are extremely delighted to announce the release of our latest software version, MSS 4.9. The Highlights of the new software include...
-
Jani Timonen created an article,
Why the trending feature does not show any values?
Sometimes even if the patient has multiple studies trending view does not show any results. Reason for the problem can be that there is no variables selected for trending. Variable selections are c...
-
Jani Timonen created an article,
Why can't MSS be installed on a Server and used by multiple users at the same time?
Info! This article is aimed for IT personnel. Medikro Spirometry Software is not designed to be used in shared server machines. MSS runs in local server machine ok. Running Medikro Spirometry Soft...
-
Jani Timonen created an article,
How can MSS run under Citrix?
Info! This article is aimed for IT personnel. Running Medikro Spirometry Software in a virtualized desktop (Citrix or Windows Metaframe) will cause several error sitautions. Mainly because spiromet...
-
Jani Timonen created an article,
How can we select the best curve in MSS?
Medikro Spirometry Software automatically select the best curve and best numeric results according ATS/ERS 2019 spirometry standard. There is no need to select best curve or numeric results manuall...
-
Jani Timonen created an article,
What reference set/predicted values should we use?
Follow your local recommendations and ask more from your local authorities. Have more questions? Submit your request here.
-
Jani Timonen created an article,
Why is the activation dialog appearing even when the spirometer is connected?
In some situations Software activation window might appear while a study is opened in MSS versions 4.5-4.7 . In MSS versions 4.5-4.7 activation logic has a timing issue, which may appear when open...
-
Jani Timonen created an article,
I get error message “Error extracting support files: Catastrophic failure”.
Info! This article is aimed for IT personnel. Most likely the reason for this is corruption in Measurements and Results (MAR) installation or uninstallation process. Solution for this is to unins...